Pular para o conteúdo principal
GE Healthcare · CTHiSpeed1100211080395651128311004

HiSpeed CTi 6.X with Octane and HSA Gantry

Symptom

Cannot network / transfer images. Cannot network / transfer images

Error message

Error: 11002 DCM low level error: System error (502): No route to host (148) See the Long Description for the full error code Error: 11002 DCM low level error: System error (502): No route to host (148) Thu Dec 29 23:00:06 2005 Error: 11002 Host: wct1 Process: dcserver File: dcscommon.c Line: 225 DCM low level error: System error (502): No route to host (148)

Error: 11080 Connection Error See the Long Description for the full error code as seen in gesyslog Thu Dec 29 23:00:01 2005 Error: 11080 Host: wct1 Process: dcserver File: dcscommon.c Line: 242 Connection error: 39565 (EFILM-WHRA) as seen in DCSLOG MESSAGE from Process 1937>> Fri Apr 7 19:00:01 2006 (dcscommon.c:242) Connection error: 11283 (PACS_Live)

Error: 11004 Failure of DCM assoc action: DCM OPEN REQ for C-STORE SCU See Long Description for Details Tue Oct 28 10:58:53 2008 Error: 11004 Host: km mod ct 01 Process: dcserver File: dcsclient.c Line: 1138 Failure of DCM assoc action: DCM OPEN REQ for C-STORE SCU

Applies to

Models
HiSpeed
Product
HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.