Symptom

Table will not move down; with tape or touch switch error See Long Description for switch locations. Table will not move down; tape switch was either contacted or not present The VCT table has a large number of touch sensors that can produce a variety of error codes.

Error message

Error: 260134717 Elevation touch switch1 (tape switches on underside of top front cover) isn''t present. Open Long Description to see error example. Tue Sep 13 13:27:51 2005 Error: 260134717 Host: tgp Process: tgp_subsys File: gtcb_emc_hardware.cxx Line: 753 Tue Sep 13 13:27:50 2005 Host : Table Sub-System Ermes # : 260134717 Exception Class : Secondary Severity : Pri/Soft File : gtcb_emc_hardware.cxx Line# : 753 Function : No System Function Reported Scan Type : None/Unknown Scan: 2081/2/0 Elevation touch switch1 (tape switches on underside of top front cover) isn''t present. Position is 840961 um from the table base.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5114124 Cable harness GTCBJ12 TO J201. see long description 5114124 Cable harness GTCBJ12 TO J201. Part #: 5114124 Part Type: GP CONSUMABLES Part Weight: 0.00 KGM Repairable: No Returnable: No Hazardous Material: N Description: GTCBJ12 TO J201 Positioning GT FRU: Yes Status: Orderable Parts Offered as: Current: 5114124 Harvested: Not Available Exchange: Not Available Comments: source pole/planner code revised per to Global Table transfer
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.