GE Healthcare · CTBrightSpeed26011819226011818426011821550368
BrightSpeed Select or Delight Select
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260118192 (50-0211)Jedi detected open filament circuit. See Long description for full error Host: Jedi Ermes #: 260118192 Exception Class: Abort Severity: Primary File: gen_exec.c Line#: 2148 Function: No System Function Reported Scan Type: Helical Scan: 542/2/1 (50-0211) Jedi detected open filament circuit. Verify tube, HV cable, HV tank filament transformer, and Jedi interconnect cables are properly installed. RMS filament current measurement (every 0.5 msec) on heater board too high. Z6,Z7= 97h,0h Error: 260118184 (50-0203) Heater inverter permanent overcurrent.(SW limit) See LD for error explanation Error: 260118184 (50-0203) Heater inverter permanent overcurrent.(SW limit) Code 50 Filament Heater Errors Error code: 50-0203H Message/explanation: Heater inverter permanent over current.(SW limit) Issued by the heater board when an over current has been detected and 3 restarts have been tried without success within 100 ms Potential cause: Heater board Recommended action/ Troubleshooting guide: Replace Heater board Error: 260118215 (60-0318) kV did not reach 75 percent after 20ms See Long Description for full error Sat Mar 25 15:35:28 2006 Error: 260118215 Host: orp Proces
Applies to
- Models
- BrightSpeed
- Product
- BrightSpeed Select, Delight Select or MERI Select Elite-16, Edge-8, Excel-4 HP_S_G16_G_HLT HP_S_G8_G_HLT HP_S_G4_G_HLT BrightSpeed Select or Delight Select (Hivalue Lite) 4/8/16 Slice Scanner BrightSpeed Delight Select Systems started to ship in first Quarter or 2009 with AIO (All in One) Console. AIO console uses HP8400 and NO DARC or IG's. Recon time is 6 FPS. If customer orders Fluoro option system requires one IG. The 16 FPS Upgrade in Not Available for Select systems. BrightSpeed Meri Select Systems ( Pancake Das ) started to ship in 3rd quarter of 2010, 4/8/16 Slice Scanner HiValue 16 Lite −−> BrightSpeed Elite Select HiValue 8 Lite−−> BrightSpeed Edge Select HiValue 4 Lite−−> BrightSpeed Excel Select BrightSpeed Select Scope: - 3.5MHu Solarix tube - JEDI 60 dc Lite - Modified NP Table Key Features: - 16x0.625mm(16x1.25) for Lite 16 - 8x1.25/8x2.5mm for Lite 4/8 - Rotation Speed: 0.8 s - Gantry: 2040mm width, 70cm Bore - Table: 1337@ISO-140mm scannable range - Recon: 3fps (w/o IG @ 2G Memory(4/8), 4G Memory(16)), optional: 6fps(w/ 1IG ) - Power: 42KW JEDI power See the attached Service Intro Planner for details Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2361419 Programmed Heater Board (eqvt 2409049) inside Jedi Auxiliary Module 2324118 See LD for Part Details - please read the note at the bottom. Part number for Heater (Filament) Board below: Part #: 2361419 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: PROGRAMMED HEATER 6P04 FRU: Yes Status: Orderable Part #: 2409049 Part Type: GP CONSUMABLES Part Weight: 0.00 KGM Repairable: No Returnable: No Hazardous Material: N Description: FRU-PROGRAMMED HEATER V6.04 FRU: Yes Status: Orderable Parts Offered as: Current: 2409049 Harvested: Not Available Exchange: Not Available Comments: added per to ECR2088514 by sso#305010201 in Jan 15, 2010 Note: * Order 2409049 to replace 2361419.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.