Symptom
Scanning aborted with unexpected Loss Of High Voltage.
Error message
Error: 183108 KV fault detected: Cathode shoot-thru. See long description for full error Sun Jun 5 23:17:05 2005 Error: 183108 Host: OBC Process: Generator File: gen_abort.c 1.31 Line: 248 Function: XRAY GENERATION : KV Loop Scan: 33003/1/1 Type: Scout Exception Level: Pri/Most Severe Time: 23:17:05:545 Log Series: 424 KV fault detected: Cathode shoot-thru. Fault latch address = ffcff3 hex, Bit = D3
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 2.X (Plus, QX/i) - GRE always an SDAS 4 slice H2_P_S4 (SDAS4 and H2 gantry) Lightspeed 2.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 3.X and 4.X. Always has an SDAS 4 slice Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-170021P92 SEMICONDUCTOR FUSE - 60A 700V Part #: 46-170021P92 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: SEMICONDUCTOR FUSE FRU: Yes Status: Orderable Comments: FLS - CLASS 1 FRU INACTIVE TO MFG. CJB 1/94 60A 0300V LIMTD RANGE SLO W UL/C DUAL ELEMENT. (REVIEWED TK, 7/91 - OK FUS RAP 60A 700V FBP60 FLS - FE ASSIST: IN ORDER TO INCREAS INCREASE REVENUE GLOBAL PARTS HAS CREATED A PROGRAM TO HARVEST PARTS FROM RENEWABLE RESOURCRESOURCES. THESE PARTS OFFER DIRECT SAVINGS TO THE BUSINESSAND THE FIELD. ORDER THE U PART WHEN AVAILABLE. IF NO U USED PARTS ARE AVAILABLE CARES WILL CHANGE THE ORDER TOA NEW PART. SEMICONDUCTOR FUSE FLS - CLASS 1 FRU INACTIVE TO MFG. CJB 1/94 60A 0300V LIMTD RANGE SLO W UL/C DUAL ELEMENT. (REVIEWED TK, 7/91 - OK FUS RAP 60A 700V FBP60
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.