GE Healthcare · CTLightSpeed2601010493024926000100726010800050052
LightSpeed 5.X (80/100kW)
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260101049 Communications with the CCB sub-system has not been established. See long description for complete error Wed Nov 22 08:29:58 2006 Error: 260101049 Host: orp Process: orp_subsys File: rcib_misc.c Line: 275 Wed Nov 22 08:29:57 2006 Host : On-Board Rotating Processor Ermes # : 260101049 Exception Class : Abort Severity : Primary File : rcib_misc.c Line# : 275 Function : No System Function Reported Scan Type : Scout Scan: 30249/1/1 Communications with the CCB sub-system has not been established. Scanning aborted. Error: 260001007 CAN serial link error. A CC error was detected when sending a message to the CCB See the long description for the full error code Tue Nov 13 21:06:38 2007 Error: 260001007 Host: OBC Process: HOP File: hop_cc_send.c Line: 196 Function: No System Function Reported Scan: 2552/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 21:06:38:745 Log Series: 937 CAN serial link error. An CC error was detected when sending a message to the CCB. Message code: 1004 Node ID: 52 Object ID: 0 Send status: 5 Error: 260108000 Collimator firmware object: Lost communications with controller GCAN error sending message = 5. See Long Description. Thu Dec 22 16:
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2159578-3 collimator control board used on 1X/2X/3X/4X/5X See Long description for details Part #: 2159578-3 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: COLLIMATOR CONTROL BOARD FRU: Yes Status: Orderable Comments: Not Available
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required