Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260118208 (60-0310) kV MAX detected See LD for error explanation Code 60 Exposure Errors Error code: 60-0310H Message/explanation: kV max detected kV reached maximum kV limit during exposure Potential cause: HV tank PPC kV control board HV cables Recommended action/ Troubleshooting guide: Repeat the exposure. Run Noload kv diagnostic without HV cables. If error persists, replace PPC kV control or HV tank.

Applies to

Models
LightSpeed, VCT, BrightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.