Pular para o conteúdo principal
GE Healthcare · CT186357

CT/e, CT Prospeed AI/FI Irix or Linux

Symptom

Cradle will not drive in or out. Cradle will not drive in or out

Error message

Error: 186357 Cradle slip was detected in response to CRADLE CMD Code: 12-1003-03 See the long description for the full error code Error: 186357 Cradle slip was detected in response to CRADLE CMD Cradle Error. Code: 12-1003-03 Error detected by: TGP (TP) Recommendation: ERRORS DETECTED BY TGP BOARD (FOR SYSTEM VERSION 5.5 OR LATER) Cradle Req Error: 3 Code: 12-1003-03 Name: Cradle Slip. Class: Error Description: Cradle slip was detected in response to CRADLE CMD. Potential Cause / FRU: 1.Patient movement during cradle movement or some obstacles on bottom of cradle or cradle roller. 2.Cradle Encoder

Applies to

Product
CT/e, CT Prospeed AI/FI Irix or Linux See long description for sale information Product Name: CT/e Single CT Scanner: Irix Platform A list of types. (a Product/hardware table) N/A An On-site / Remote identification aid . (Example: check_config) N/A A Table of options (included, optional, new introductions, …) N/A A list of Software that has run on this product (Product / Software Compatibility Table) 3.01 / 3.02 / 3.03 Linux CTe Product photographs N/A Marketing Information N/A Documents: Always use the most appropriate documentation for your system. Documentation can be found on CT Service CDROM/DVD, or online at FCT Service Engineering Tech Pubs Website, Common Document Library, and My Workshop. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found in the attachment Always use the most appropriate documentation for your system.
Component
2299288 - STEP MOTOR DRIVER ASSEMBLY See long description for details Part #: 2299288 Part Type: GP CONSUMABLES Part Weight: 0.50 KGM Repairable: No Returnable: No Hazardous Material: N Description: DRIVER FRU: Yes Status: Orderable Parts Offered as: Current: 2299288 Comments: FLS - STEP MOTOR DRIVER ASSEMBLY ADDED PER RSPL FOR LOCALIZED CT/E TABLE MSA8429 10/24/01
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues