GE Healthcare · CTLightSpeed260107236
LightSpeed 5.X (80/100kW)
Symptom
"Scan hardware Stopped Scan" with Axial Errors. "Scan hardware Stopped Scan" with Axial Errors
Error message
Error: 260107236 Axial Motor Control (AMC) Logic will NOT Comand the DRIVE until the Gantry motion is STOPPED. See Long Description for full error Mon May 2 09:25:58 2005 Error: 260107236 Host: tgp Process: tgp_subsys File: amc_mgr.c Line: 2385 Mon May 2 09:25:58 2005 Host : Table Gantry Processor Ermes # : 260107236 Exception Class : Abort Severity : Primary File : amc_mgr.c Line# : 2385 Function : AXIAL CONTROL : Axial Control Scan Type : Helical Scan: 19/1/1 Axial Motor Control (AMC) Logic will NOT Comand the DRIVE until the Gantry motion is STOPPED. The Encoder and the Drive MUST be Sync'd/Referenced. (Cur State: 10 Cmd: 21) HW/SW Error: Retry Rx to verify probable Gantry Hardware problem.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5120644 Cable - Ax Drive to MSUB J29 - EMC Editi See Long Description for part details. Part #: 5120644 Part Type: GP CONSUMABLES Part Weight: 0.30 KGM Repairable: No Returnable: No Hazardous Material: N Description: Cable - Ax Drive to MSUB J29 - EMC Editi FRU: Yes Status: Orderable Comments: added per VCT rspl, this part is EMC2 Compliant, ms13452 041107
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.