Symptom

Cannot scan with Data Acquisition System Temperature is nearing Upper Limit

Error message

Error: 260002470 Highest temperature was seen is on Converter Board 41. See LD for more details Mon Apr 10 08:21:33 2006 Error: 260002470 Host: DAS Process: DCB Firmware File: ctemptst.c Line: 800 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 50346/8/39 Type: Axial Exception Level: Pri/Soft Time: 08:21:33:105 Log Series: 32 WARNING: DAS Temperature is nearing the Upper Limit. Highest temperature was seen is on Converter Board 41. Measured: 55.5 degrees C. Acceptable Range: 15.0 to 62.0 degrees C. Converter Board Bit Map: 0x010000000000. Bit Map Indicates Failed Board(s) for Multiple Board Errors. (Board Bit Map defines: BIT47 = CnvBrd 48 ... BIT0 = CnvBrd 1)

Applies to

Models
LightSpeed
Product
LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2125040 AIR FLW: 79.4CFM, 115V, 60HZ See long description for details 2125040 AIR FLW: 79.4CFM, 115V, 60HZ Part #: 2125040 Part Type: GP CONSUMABLES Part Weight: 0.91 LBR Repairable: No Returnable: No Hazardous Material: N Description: AIR FLW: 79.4CFM, 115V, 60HZ FRU: Yes Status: Orderable Comments: FLS - ADDED BY PCN 189406. MS4636 7/18/95 GELS - "IT'S A 115V,60HZ FAN ./.",FAN 119X119 AIR FLW:79.4CFM 115V
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.