Lightspeed 1.X - Octane
Symptom
Scan Hardware Stopped Scan with DCB has an Outstanding SelfTest Failure See Long Description for error string Scan Hardware Stopped Scan message. The DCB has an Outstanding SelfTest Failure which does NOT allow Patient Scanning. Data Acquisition System Has Detected An Error In The 5V Analog DAS Power Supply, and DCB Power Supply selftest failed Errors posted in gesyslog whenever a tube warm up was attempted: Tue Dec 7 07:23:22 2004 Error: 245043 Host: ct02_sbc Process: scanmgr File: scanmgr_events.cxx Line: 1595 Function: System Control : Scan Control Scan Type: Cine Scan: 0/0/0 Scan Seq Id: 1000 Scanning System Aborted Scanning Tue Dec 7 07:23:22 2004 Error: 260002455 Host: DAS Process: DCB Firmware File: scanrx.c Line: 290 Function: DATA ACQUSITION : Scan Data Acquisition Scan: 50723/1/1 Type: Axial Exception Level: Pri/Soft Time: 07:23:22:900 Log Series: 36 The DCB has an Outstanding SelfTest Failure which does NOT allow Patient Scanning. (The DCB current State: Service State ). Tue Dec 7 07:23:22 2004 Error: 260001023 Host: OBC Process: HOP File: hop_erlog.c Line: 367 Function: No System Function Reported Scan: 50723/1/1 Type: Axial Exception Level: Pri/Most Severe Time: 07:23:22:910 Log Series: 37 The DCB has notified the system of an error condition which prevents scanning. See log for further information:^C Error message code: 260002358 Notification level: 5 Sequence number: 4 Tue Dec 7 07:23:22 2004 Error: 260002358 Host: DAS Process: DCB Firmware File: scanrx.c Line: 677 Function: DATA ACQUSITION : Scan Data Acquisition Scan: 50723/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 07:23:23:210 Log Series: 37 Scan ABORT -- DAS Has Detected An Error.^B Scan Setup Errors Prevent The Data Acquisition System From Being Able To Scan. (See Previous Log Entries.) Errors noted previously in the geyslog, prior to the customer attempting a tube warm up. These messages occurred during the prior system boot up: Tue Dec 7 07:20:51 2004 Error: 260001023 Host: OBC Process: HOP File: hop_erlog.c Line: 367 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 07:20:50:115 Log Series: 23 The DCB has notified the system of an error condition which prevents scanning. See log for further information:^C Error message code: 260002403 Notification level: 5 Sequence number: 1 Tue Dec 7 07:20:51 2004 Error: 260002403 Host: DAS Process: DCB Firmware File: pwrtst.c Line: 729 Function: DATA ACQUSITION : Scan Data Acquisition Scan: 0/0/0 Type: None/Unknown Exception Level: Sec/Soft Time: 07:20:50:140 Log Series: 23 The Data Acquisition System Has Detected An Error In The +5V Analog DAS Power Supply.^B Measured: 2522 mV, Expected range: 4750 to 5250 mV Tue Dec 7 07:20:51 2004 Error: 260002403 Host: DAS Process: DCB Firmware File: pwrtst.c Line: 783 Function: DATA ACQUSITION : Scan Data Acquisition Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 07:20:50:240 Log Series: 23 The Data Acquisition System Has Detected An Error In The -5V Analog DAS Power Supply.^B Measured: 508 mV, Expected range: -5250 to -4750 mV Tue Dec 7 07:20:51 2004 Error: 260002318 Host: DAS Process: DCB Firmware File: hdwtest.c Line: 110 Function: DATA ACQUSITION : Scan Data Acquisition Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 07:20:50:505 Log Series: 23 A Failure Has Been Detected By A DCB Selftest.^B DCB Power Supply selftest failed.
Error message
Error: 260002455 The DCB has an Outstanding SelfTest Failure which does NOT allow Patient Scanning. (The DCB current State: Service State ) See Long Description for the full error code Tue Dec 7 07:23:22 2004 Error: 260002455 Host: DAS Process: DCB Firmware File: scanrx.c Line: 290 Function: DATA ACQUSITION : Scan Data Acquisition Scan: 50723/1/1 Type: Axial Exception Level: Pri/Soft Time: 07:23:22:900 Log Series: 36 The DCB has an Outstanding SelfTest Failure which does NOT allow Patient Scanning. (The DCB current State: Service State ).
Applies to
- Product
- Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2218158 SDAS +/-5VDC ANALOG PWR SUPPLY W/BRACKET. used on 1.x with H1 gantry See Long Description for details Part #: 2218158 Part Type: GP CONSUMABLES Part Weight: 1.81 LBR Repairable: No Returnable: No Hazardous Material: N Description: DAS 5VDC ANALOG PWR SUPPLY W/BRACKET FRU: Yes Status: Orderable Comments: FLS - DAS +/-5VDC ANALOG PWR SUPPLY W/BRACKET U/O HELIOS DAS REPLACES 46-251198P46 MS08992 10/5/98 GENCO INDUSTRIES-WAUKESHA,WI PER FE - 46-251198P46 IS NOT ENTIRELY COMPATIBLE - 3 CONNECTION POINTS FOR WIRES - NEED 4- CANNOT SOLDER WIRE IN CORRECTLY DUE TO HOLE IN SOLDER POINT BEING TOO SMALL FOR MORE THAN 1 WIRE. TW/01/03
- Area
- CT