Pular para o conteúdo principal
GE Healthcare · CTLightSpeed230017082394680230017078503421111522307108580

LightSpeed 2.X (Plus, QX/i) - Octane

Symptom

Fastcal is failing with beam obstructed errors. Fastcal is failing with beam obstructed errors

Error message

Error: 230017082 Channel in non-bowtie airscan out of range - row 0, channel 10, value = 4.3 The row number can be VERY confusing. Please see Long Description for how to decode the ROW NUMBER Tue Mar 22 20:12:04 2005 Error: 230017082 Host: ct01 Process: accs File: aircal_processing.c Line: 917 Channel in non-bowtie airscan out of range - row 0, channel 10, value = 4.394680 Click Here to Decode the ROW NUMBER

Error: 230017078 Aircal failed to process scan See Long Description for complete error Tue Mar 22 20:12:04 2005 Error: 230017078 Host: ct01 Process: accs File: accs_events.cxx Line: 463 Aircal failed to process scan - Exam = 50342, Series = 1, Scan = 1, Seconds = 1111522307 Microseconds = 108580 View scanfileto check data and header information

Error: 230023046 Calibration limit check failed. Beam may be obstructed See LD for complete error Tue Mar 22 20:12:04 2005 Error: 230023046 Host: ct01 Process: dailyPrepRx File: DailyPrepUtils.c Line: 3953 Calibration limit check failed. Beam may be obstructed

Applies to

Models
LightSpeed
Product
LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.