GE Healthcare · CTLightSpeed230017821470544300000230017801
LightSpeed 4.X - GRE
Symptom
Collimator calibration drift check has failed Cal will continue with the next step - FPA Check.
Error message
Error: 230017821 ACCS CCal: Drift check: The polynomial value has drifted by 0.470544 See the LD for the complete error code Error: 230017821 ACCS CCal: Drift check: The polynomial value has drifted Wed Aug 9 08:18:38 2006 Error: 230017821 Host: qxi1_sbc Process: accs File: collSweepProcessing.c Line: 1635 ACCS CCal: Drift check: The polynomial value has drifted by 0.470544 which is more than tolerance 0.300000 Error: 230017801 ACCS CCal: Sweep Processing failed. Status = -1 See the Long Description for the full error code Tue Nov 30 02:33:04 2004 Error: 230017801 Host: k2_ct2_ls Process: accs File: ccal_prep.c Line: 661 ACCS CCal: Sweep Processing failed. Status = -1
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2318808-3 150 PIN - 5 connector INTERPOSER Replaces part 2318808-2 used in MDAS-16 and GDAS-16 (Quantity = 5) 2318808-3 150 PIN - 5 connector INTERPOSER Replaces part 2318808-2 used in MDAS-16 and GDAS-16 systems. Part #: 2318808-3 Part Type: GP CONSUMABLES Part Weight: 1.09 LBR Repairable: No Returnable: No Hazardous Material: N Description: (5) CONNECTORS, 150 PIN INTERPOSER FRU: Yes Status: Orderable Comments: FLS - CONNECTORS (5). ADDED PER CT PCN 220814. REPLACES P/N 2318808-2 SINCE THIS -2 PART CHANGE FROM FRU YES TO NO. MS20544 5/20/03 PCN NOTES: 2318808-3 IS A TUBE THAT CONTAINS 5 PIECES. FOR QUANTITY 1 ORDERED FE WILL RECEIVE 5 INTERPOSERS. VENDOR INFO MILL-MAX MFG. CORP S596-00-000-00-955000 VC 462905 8536.69.4040 HTS CODE
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.