GE Healthcare · CTLightSpeed18414217231184188170851713517140
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane
Symptom
Missed ISD See Long Description Inter Scan Delay ISD Missed .
Error message
Error: 184142 ISD Missed See Long description for complete error Wed Nov 8 11:01:40 2006 Error: 184142 Host: STC Process: Scan Control File: scTrigSeqActions.c 1.65 Line: 1043 Function: No System Function Reported Scan: 496/2/2 Type: Axial Exception Level: Pri/Soft Time: 11:01:40:295 Log Series: 937 ISD missed. Rx ISD: 1000 mS Requested ISD = 1 Second Actual ISD: 17231 Actual ISD = 17 Seconds More Data Follows: See Error: 184188 for additional detail Error: 184188 Timing (in mS) since end of last scan Col:17085 DvRdy:17135 GenX:17135 TrigEn:17140 ExpEn:17230 See Long Description for more details Wed Nov 8 11:01:40 2006 Error: 184188 Host: STC Process: Scan Control File: scTrigSeqActions.c 1.65 Line: 1048 Function: No System Function Reported Scan: 496/2/2 Type: Axial Exception Level: Pri/Soft Time: 11:01:40:295 Log Series: 938 Last Hdw Rdys- Offsets:0x40 Cpb:0x20 Xray:0x80 Timing (in mS) since end of last scan. RxRcvd:-670 DoRdy:95 OfSt:165 Cpb:17085 OfStp:17095 Cr:810 Ax:100 GenP:175 Das:160 Col:17085 DvRdy:17135 GenX:17135 TrigEn:17140 ExpEn:17230 (Rdy:0xf00e2fff Of:0x540 Auto:0x57c Cpb:0x4057c Xray:0x80bff) This Error code is can give you a clue as to where the problem is comin
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5331736 - Helios Filter Asselmbly FRU (replaces 2291958) Used on Helios 3 and 4 Collimators. See Long Description for more Details Part #: 5331736 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Helios Filter Asselmbly FRU FRU: Yes Status: OrderableComments: Per ECR 2082045, 212008898, 6/16/09. Part #: 2126674 Part Type: GP CONSUMABLES Part Weight: -144548.55 LBR Repairable: No Returnable: No Hazardous Material: N Description: FILTER ASSEMBLY, HELIOS COLLIMATOR FRU: NO Status: Non Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.