Pular para o conteúdo principal
GE Healthcare · CTBrightSpeed11046

BrightSpeed Select or Delight Select

Symptom

Site cannot send a specific exam or series to PACS or other dicom hosts. All other exams send without error. See Long Description for more details Site cannot send a specific patient to PACS or other dicom host. All other exams send without error. Error as seen in the GESYSLOG Fri Apr 7 19:00:01 2006 Error: 11046 Host: fchct1 Process: dcserver File: dcspush.c Line: 1381 STORE response received with refused status (0xA700) - Aborting! Fri Apr 7 19:00:01 2006 Error: 11009 Host: fchct1 Process: dcserver File: dcspush.c Line: 1516 Association was aborted with PACS_Live Fri Apr 7 19:00:01 2006 Error: 11080 Host: fchct1 Process: dcserver File: dcscommon.c Line: 242 Connection error: 11283 (PACS_Live) Error as seen in the DCSLOG MESSAGE from Process 1937>> Fri Apr 7 19:00:01 2006 (dcspush.c:1381) [Dcserver]> STORE response received with refused status (0xA700) - Aborting! MESSAGE from Process 1937>> Fri Apr 7 19:00:01 2006 (dcspush.c:1516) [Dcserver]> Association was aborted with PACS_Live MESSAGE from Process 1937>> Fri Apr 7 19:00:01 2006 (dcscommon.c:242) Connection error: 11283 (PACS_Live)

Error message

Error: 11046 STORE response received with refused status (0xA700) - Aborting! See LD for full text. Error: 11046 STORE response received with refused status (0xA700) - Aborting! Error as seen in the GESYSLOG Fri Apr 7 19:00:01 2006 Error: 11046 Host: fchct1 Process: dcserver File: dcspush.c Line: 1381 STORE response received with refused status (0xA700) - Aborting! Error as seen in the DCSLOG MESSAGE from Process 1937>> Fri Apr 7 19:00:01 2006 (dcspush.c:1381) [Dcserver]> STORE response received with refused status (0xA700) - Aborting!

Applies to

Models
BrightSpeed
Product
BrightSpeed Select, Delight Select or MERI Select Elite-16, Edge-8, Excel-4 HP_S_G16_G_HLT HP_S_G8_G_HLT HP_S_G4_G_HLT BrightSpeed Select or Delight Select (Hivalue Lite) 4/8/16 Slice Scanner BrightSpeed Delight Select Systems started to ship in first Quarter or 2009 with AIO (All in One) Console. AIO console uses HP8400 and NO DARC or IG's. Recon time is 6 FPS. If customer orders Fluoro option system requires one IG. The 16 FPS Upgrade in Not Available for Select systems. BrightSpeed Meri Select Systems ( Pancake Das ) started to ship in 3rd quarter of 2010, 4/8/16 Slice Scanner HiValue 16 Lite −−> BrightSpeed Elite Select HiValue 8 Lite−−> BrightSpeed Edge Select HiValue 4 Lite−−> BrightSpeed Excel Select BrightSpeed Select Scope: - 3.5MHu Solarix tube - JEDI 60 dc Lite - Modified NP Table Key Features: - 16x0.625mm(16x1.25) for Lite 16 - 8x1.25/8x2.5mm for Lite 4/8 - Rotation Speed: 0.8 s - Gantry: 2040mm width, 70cm Bore - Table: 1337@ISO-140mm scannable range - Recon: 3fps (w/o IG @ 2G Memory(4/8), 4G Memory(16)), optional: 6fps(w/ 1IG ) - Power: 42KW JEDI power See the attached Service Intro Planner for details Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues