Pular para o conteúdo principal
GE Healthcare · CTHiSpeed183113

HiSpeed QX/i - Octane

Symptom

No flashing eights on gantry display and Scanning hardware reset failed

Error message

Error: 183113 KV board HVON bit high on power-up. Fri Feb 17 14:31:54 2006 Error: 183113 Host: OBC Process: Control File: Line: 235 Function: XRAY GENERATION : KV Loop Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 1067 Log Series: 2 KV board HVON bit high on power-up. Hardware address = 0xffcff9, Status = 0x81

Applies to

Models
HiSpeed
Product
HiSpeed QX/i - Octane Always MDAS4 101I.2_HSQXIM5 101.4_HSQXIM4 HiSpeed QX/i - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 101I.2_HSQXIM5 101.4_HSQXIM4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2143147 KV BOARD T3202HK 2143147 KV BOARD T3202HK 2143147 KV BOARD Part #: 2143147 Part Type: GP REPAIRABLES Part Weight: 1.81 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: KV BOARD FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.