Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT

LightSpeed 1.X - GRE

Symptom

A Pink Pop up message is reported on boot-up that Recon times may be slower - scanning may continue See Long description to see the Pink Pop up message A Pink Pop up message is reported on boot-up that Recon times may be slower

Error message

DARC /var/log/messages : timeout attempting to rsh to ig2 ( can be ig1 or ig3 ) See long description DARC /var/log/messages Aug 18 18:43:09 darc logger: is_IG_up 1 ig2 called Aug 18 18:43:09 darc is_IG_up:[recon_control:7973]: ig3 is up ( this will be typically 5 mins after start of DARC boot ) Aug 18 18:43:09 darc is_IG_up:[recon_control:7967]: ig1 is up Aug 18 18:43:10 darc is_IG_up:[recon_control:7970]: timeout attempting to rsh to ig2 ( this is ig2 failing to respond to DARC ) With no intervention DARC will retry to contact ig that failed to respond on fisrt attempt ( notice timestamps ) Aug 18 18:49:58 darc logger: is_IG_up 3 ig2 called Aug 18 18:49:58 darc mountd[3273]: authenticated mount request from ig2:855 for /usr/g (/usr/g) Aug 18 18:49:58 darc is_IG_up:[recon_control:7970]: ig2 is up ( ig2 up nearly 7 minutes after second start )

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 1.X - GRE H1_P_S4 Lightspeed 1.X - GRE Method of visual identification Large square gantry similar in appearance to a CTi gantry. Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required