Symptom

Scan aborts. Scans abort

Error message

Error: 260118184 (50-0203) Heater inverter permanent overcurrent.(SW limit) See LD for error explanation Error: 260118184 (50-0203) Heater inverter permanent overcurrent.(SW limit) Code 50 Filament Heater Errors Error code: 50-0203H Message/explanation: Heater inverter permanent over current.(SW limit) Issued by the heater board when an over current has been detected and 3 restarts have been tried without success within 100 ms Potential cause: Heater board Recommended action/ Troubleshooting guide: Replace Heater board

Error: 260118192 (50-0211)Jedi detected open filament circuit. See Long description for full error Host: Jedi Ermes #: 260118192 Exception Class: Abort Severity: Primary File: gen_exec.c Line#: 2148 Function: No System Function Reported Scan Type: Helical Scan: 542/2/1 (50-0211) Jedi detected open filament circuit. Verify tube, HV cable, HV tank filament transformer, and Jedi interconnect cables are properly installed. RMS filament current measurement (every 0.5 msec) on heater board too high. Z6,Z7= 97h,0h

Applies to

Models
BrightSpeed, VCT
Product
BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.