Pular para o conteúdo principal
GE Healthcare · CTLightSpeed184836182405100642072920807

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 184836 motor controller detects uncommanded motion. Aborting any active scan Tue Aug 10 19:04:09 2004 Error: 184836 Host: ETC Process: Elevation File: elev scan.c 1.17 9/30/92 Line: 178 Function: No System Function Reported Scan: 4530/2/7 Type: Helical Exception Level: Pri/Most Severe Time: 19:04:08:905 Log Series: 376 motor controller detects uncommanded motion. Aborting any active scan

Error: 182405 HW/SW error: Uncommanded elevation motion. See long description for full error Wed Mar 19 13:22:41 2008 Error: 182405 Host: ETC Process: Elevation File: unComMotion.c 1.28 11/5/92 Line: 170 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 10064/2/6 Type: Axial Exception Level: Pri/Soft Time: 13:22:42:215 Log Series: 54 HW/SW error: Uncommanded elevation motion. Current Position 20729 (.01 mm) Last Stationary Position 20807 (.01 mm) Position Tolerance 75 (.01 mm) Current Encoder cnts 3446 Last Stationary Encoder Cnts 3435

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2100671 CT TABLE GAS SPRING - Used on ALL HSA, CTi, LightSpeed 1.X through 5.X tables Part #: 2100671 Part Type: GP CONSUMABLES Part Weight: 1.40 LBR Repairable: No Returnable: No Hazardous Material: N Description: CT Table gas spring FRU: Yes Status: Orderable Comments : FLS - REPLACES 46-327064P1 & 46-296891P1 PER PCN 161843 7-20-93 MS4398. NOT ON SSS: SUSPA IS NOT SUPPLIER. TOOL FOR REPLACING GAS SPRING IS 2144261 ************************************************************PART HAS BEEN DE-LISTED AS A HAZARDOUS MATERIAL ITEM. PART IS NOT HAZARDOUS FOR ANY MODE OF TRANSPORTATION. HAZARDOUS PAPERWORK IS NOT NEEDED. 4/99 ************************************************************ ALSO USED ON QXI LIGHTSPEED ALSO USED ON LIGHTSPEED GELS - "NON FLAMMABLE GAS - HAZ IATA DESC. ""ARTICLES, PRESSURIZED, PNEUMATIC "" UN 3164 - CLASS 2.2",GAS SPRING HAZ #8 is the shock
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.