Pular para o conteúdo principal
GE Healthcare · CTLightSpeed185109178574178568

LightSpeed 4.X - GRE

Symptom

Table will not move. Table will not move

Error message

Error: 185109 Elevation Interference/Limit switches are INVALID. See Long Description for full error Mon Dec 10 16:56:08 2007 Error: 185109 Host: ETC Process: Control File: etcValidIntfMatrix.c 1.8 7/9/ Line: 96 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 16:56:08:425 Log Series: 45 HW error: Elevation Interference/Limit switches are INVALID. Interference/Limit switches = 0xc1d0. UP LIMIT: 1 DOWN LIMIT: 0 INTF1: 1 INTF2: 0 (0 = NOT active 1 = active)

Error: 178574 The following controllers STC/ETC/OBC failed to report initialization status in 120 seconds See Long description for error code variations Sun Nov 27 14:18:24 2005 Error: 178568 Suite: SBCT Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to ETC in 10 seconds ------------------------------------------------------------- Wed Sep 8 06:20:02 2004 Error: 178568 Host: cmls Process: fwmgr File: NBSSocket.c Line: 388 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failed to establish socket connection to OBCR in 10 seconds. -------------------------------

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues