GE Healthcare · CTHiSpeed980042450681771281001
System Does not Scan with Raw Data Error
Symptom
System Does not Scan with Raw Data Error System Does not Scan with error Raw Data Error
Error message
Error: 98004 Failed to read 2048 bytes from raw partition /dev/rsd2h See long description for complete error. Mon Jan 31 14:45:09 2005 Error: 98004 Suite: OCT2 Host: SBC0 Process: dataacq File: RPDiskSCSIWithCache.m 1.11 Method: devRead:length: Line: 654 Failed to read 2048 bytes from raw partition /dev/rsd2h, count=-1, errno=5 Error: 245068 : Continuous Scan Status Message: The System Aborted Scanning:, views expected 17712 - views collected 0 See Long Description for complete error. Error: 245068 : Continuous Scan Status Message: The System Aborted Scanning:, views expected 17712 - views collected 0 Tue Nov 24 13:08:44 2009 Error: 245068 Suite: CT01 Host: SBC0 Process:Scanmgr File: do_continuous_complete.m 1.44 Method: do_continuous_complete() Line: 360 Function: Data Acquisition : SBCProcessing Scan: 6011/2/2 Type: Helical Scan Seq: 81001 Continuous Scan Status Message: The System Aborted Scanning:, views expected 17712 - views collected 0
Applies to
- Models
- HiSpeed
- Product
- HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2226715-3 CTI 18GB ST318418N (replaces 2226715-2 CTI 9 GB) Scan Data Disk, Used on Indigo OC, sbc local scan disk and sbc scan data disk -and Octane scan data disk See LD for more information and Picture Part#: 2226715-3 replaces 2226715-2 ( also shown below for reference ) Part #: 2226715-3 Part Type: GP CONSUMABLES Part Weight: -98999.01 LBR Repairable: No Returnable: No Hazardous Material: N Description: HARD DRIVE 18GB (DSTRK 9GB) FRU: Yes Status: Orderable Comments: *** OBSOLETE PART CASE 5523893 *** This part has been identified as obsolete and no short-term solution has been identified. FE must identify and order a higher or lower level spare part as required. Contact your support engineer, on-line engineer or Parts ID team should you require support. 212012173 10-2-06 11/19/04 (Shawn Fox 212001261) Please return your defective Drive's to: Babar Maqsood 3000 N. Grandview Blvd Mailstop W-1250 Waukesha, WI 53188 FE ASSIST MS6588 MARCH 2002 2226715-3 = ST318418N 18GB DRIVE. 2226715-3 IS A DESTROKED DRIVE 2226715-3 CAN BE USED IN THE FOLLOWING LOCATION: CTI /W INDIGO -3 WORKS IN OC SBC/LSD & SCAN DATA DISK. CTI /W OCTANE -3 WORKS IN SBC/LSD & SCAN DATA DISK. WHEN USING A -3 AS A CTI INDIGO OC DISK PLEASE READ: FOR CTI INDIGO OC DISK A SOFTWARE SCRIPT IS NEEDED TO LOAD SOFTWARE. THE SOFTWARE SCRIPT CAN BE FOUND ON SUPPORTCENTRAL OR IN THE CLARIFY PROBLEM SOLUTION DATABASE. 212002185 05/10/2007 22:28 GMT CTI 9 GB Scan Data Disk, part# 2226715-2 Used on Indigo OC, sbc local scan disk and sbc scan data disk - and Octane scan data disk picture below Part #: 2226715-2 Part Type: GP CONSUMABLES Part Weight: 0.91 LBR Repairable: No Returnable: No Hazardous Material: N Description: SCAN DISK SBC SYSTEM 9.2 GB USE WITH 2204245-3 CD ROM 6.2 OS FRU: Yes Status: Orderable Comments: 11/19/04 (Shawn Fox 212001261) 2226715-2 = NEW 9 GB DRIVE SEAGATE ST39216N. 2226715-2 IS USED ON INDIGO (CTI4.4 & FORWARD) SBC LOCAL SCSI DISK AND INDIGO OC DISK. 2226715-2 IS USED ON OCTANE SYSTEMS (CTI 6.1 & FORWARD) INTHE SBC LOCAL SCSI DISK. Below is a pic of the disk location from CT Hispeed Systems FRU Manual, Direction 46-018316 Rev 37 Octane OC below
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.