GE Healthcare · CTBrightSpeed260115193260115194
BrightSpeed HPower or Delight HPower
Symptom
Table will not drive up or down
Error message
Error: 260115193 Elevation drive fault occurred. Bus over voltage detected by DSP, External Regen Fault. See long description Tue Nov 3 08:04:58 2009 Error: 260115193 Host: tgp Process: tgp_subsys File: mtcb_emc_action.c Line: 213 Tue Nov 3 08:04:57 2009 Host : Table Sub-System Ermes # : 260115193 Exception Class : Secondary Severity : Pri/Soft File : mtcb_emc_action.c Line# : 213 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Elevation drive fault occurred. FaultCode : Bus over voltage detected by DSP, External Regen Fault ExtFault : Not Assigned Sat Jun 12 18:09:51 2010 Error: 260115193 Host: tgp Process: tgp_subsys File: mtcb_emc_action.c Line: 213 Sat Jun 12 18:09:52 2010 Host : Table Sub-System Ermes # : 260115193 Exception Class : Soft Severity : Pri/Soft File : mtcb_emc_action.c Line# : 213 Function : No System Function Reported Scan Type : None/Unknown Scan: 6448/2/5 Elevation drive fault occurred. FaultCode : Resolver Fault ExtFault : Not Assigned Error: 260115194 Elevation drive fatal fault occurred. See long description Tue Nov 3 08:04:58 2009 Error: 260115194 Host: tgp Process: tgp_subsys File: mtcb_emc_action.c Line: 206 Tue Nov 3 08:04:57
Applies to
- Models
- BrightSpeed
- Product
- BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2342729-3 Elevation or Cradle Drive Amplifier used on the 5.X table - same driver amp is used for both See LD for additional part information Part #: 2342729-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: H-POWER TABLE SERVO DRIVE, EMC EDITION 2 COMPLIANT FRU: Yes Status: Orderable Comments: **Please consider ordering the used part when it is appropriate for your customer. It will save you substantial money.** Part added per DST-HP60 16 slice RSPL. 040906 MS13452.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required