Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT230015028230017089230017078503421111522307108580

LightSpeed 1.X - GRE

Symptom

DAILY IQ CHECK THRESHOLD FAILURE

Error message

Error: 230015028 Failed to construct a file path for vector type 2 See long description for complete error Fri Feb 2 06:16:14 2007 Error: 230015028 Host: ls16 Process: accs File: cal db api.c Line: 1893 Function: CAL Processing : SBC Processing Failed to construct a file path for vector type 2 in directory /usr/g/cal db/Value:48/large sfov/16x0625/small spot/dasgain2 .

Error: 230017089 Couldn't write acal vector into cal database See long description for full error Thu Jun 5 07:00:23 2008 Error: 230017089 Host: ct01 Process: accs File: aircal_processing.c Line: 1334 Couldn't write acal vector into cal database

Error: 230017078 Aircal failed to process scan See Long Description for complete error Tue Mar 22 20:12:04 2005 Error: 230017078 Host: ct01 Process: accs File: accs_events.cxx Line: 463 Aircal failed to process scan - Exam = 50342, Series = 1, Scan = 1, Seconds = 1111522307 Microseconds = 108580 View scanfileto check data and header information

Error: 230017086 DAILY IQ CHECK THRESHOLD FAILURE See Long Description for the full error Error: 230017086 Host: mcop_sbc Process: accs File: aircal_processing.c Line: 797 DAILY IQ CHECK THRESHOLD FAILURE - Scan 50055/8/3 Drift chec

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 1.X - GRE H1_P_S4 Lightspeed 1.X - GRE Method of visual identification Large square gantry similar in appearance to a CTi gantry. Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.