Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2300150355120052300150052300150241127205412245201

LightSpeed 5.X (60kW)

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 230015035 Cal_db failure See LD for complete error Tue Oct 18 07:20:43 2005 Error: 230015035 Host: ctls3 Process: scanmgr File: cal_db_file_io.c Line: 762 Cal_db failure, error number = 12, reason: Not enough space Tue Sep 20 08:36:512005 Error: 230015035 Host: ls16 Process: scanmgr File: cal_db_file_io.c Line: 231 Cal_db failure, error number = 98, reason: Address already in use

Error: 230015005 Unable to retrieve the requested cal module from the Cal Database Sun Jun 19 21:05:13 2005 Error: 230015005 Host: wodct03403 Process: scanmgr File: cal_db_api.c Line: 451 Function: CAL Processing : SBC Processing Unable to retrieve the requested cal module from the Cal Database

Error: 230015024 A corrupt file has been found in the cal database See LD for complete error Tue Sep 20 08:36:51 2005 Error: 230015024 Host: ls16 Process: scanmgr File: cal_db_file_io.c Line: 1983 Function: CAL Processing : SBC Processing A corrupt file has been found in the cal database. The corrupt file has been removed from the cal database and saved for later analysis. Corrupt File: /usr/g/cal_db/kv120/large_sfov/16x0625/large_spot/dasgain5/cal.1.acal Saved File: /usr/g/cal_db/recovery/kv120.large_sfov.

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.