Pular para o conteúdo principal
GE Healthcare · CTHiSpeed18880913592

CT/e Dual, CT ProSpeed AII/FII Irix or Linux

Symptom

System boots to applications, then shuts down Error: 200002352 HOST SHUTDOWN was initiated by the unexcpected termination of recon control.

Error message

Error: 188809 Error in _rdh_readRawHeader: RawRead See Long Description for complete error Mon Mar 13 19:15:55 2006 Error: 188809 Host: CT01 Process: RawMgr File: rawfile.c Line: 517 Error in _rdh_readRawHeader: RawRead(header,13592) = 1.

Applies to

Models
HiSpeed
Product
CT/e Dual, CT ProSpeed AII/FII Irix or Linux HiSpeed CT/e Dual CT Scanner System see long Description for sales information Product Name: CT/e Dual CT Scanner System Irix A list of types. (a Product/hardware table) N/A An On-site / Remote identification aid . (Example: check_config) N/A A Table of options (included, optional, new introductions, …) N/A A list of Software that has run on this product (Product / Software Compatibility Table) N/A Product photographs N/A Marketing Information N/A Documents: Always use the most appropriate documentation for your system. Documentation can be found on CT Service CDROM/DVD, or online at FCT Service Engineering Tech Pubs Website, Common Document Library, and My Workshop. Links: Drawing Print 2320306-2CCT : http://gemspdm.med.ge.com/ematrix/common/eLibDocument.jsp?type=type_DrawingPrint&name=2320306-2CCT&rev=9 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5166701 (replaces 2376479) SCSI HDD 36GB 3.5IN Recon NP See long description for details Part #: 5166701 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: SCSI HDD 36GB 3.5IN Recon NP FRU: Yes Status: Orderable Comments: Part added per ECR2027606 305014118 06/19/2006 Part #: 2376479 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Description: P2011US 36GB 3.5"SCSI HDD FRU: Yes Status: Orderable Comments: revised per to GEHW confirmation ** FE ASSIST ** If no stock of the 2376479 can use 5166701 per ECR 2027606. (212001261 5/9/07)
Area
CT