Symptom

Unable to scan with XG errors

Error message

ER-113 XG Error - contains information on all XG ER113 errors See long description for detailed error codes and sub codes. ER–113 XG Error LED STATUS = uvwxyz, %%%KV, %%%MA, HV TIME=%%%%%msec NV TIME: Time between HV ON and XG FAIL LED STATUS=uvwxyz: This hexadecimal code shows the LED status (D23 ~ D39) on the Power Cont Assy into the X0Ray Generator as shown: u: 0 ~ 3 0 0 D23 D24 D23: Over voltage P4 D24: Over voltage N4 v: 0 ~ F D25 D26 D27 D26 D25: Over Voltage P1 D26: Over Voltage N1 D27: Down Voltage P4 D28: Down Voltage N4 w: 0 ~ 3 0 0 D29 D30 D29: Over Load P D30: Over Load N x: 0 ~ F D31 D32 D33 D34 D31: Over Heat D32: Rotor Over Current D33: Filament Over Current D34: Power Fail y: 0 z: 1 ~ 7 D35 D36 D37 D38 1: HV ON Over time (more than 14.67sec below 100mA) 2: HV ON Over time (more than 5.64sec above 100mA) 3: FIL3 SW does not match soft SW. 4: Can not select 80 or 140 kV for soft SW. 5: RAM over error 6: RAM set error 7: RAM clear error

ER-117 XG Fail - contains information on all XG ER117 errors See Long Description ER-117 XG Fail ER–117 ER – 117 : XG Error 00000XYY Error Code X = _ 0: Test Mode 1: Stand–by 2: Preparation Acting 3: Generator Ready 5: Exposure 7: Tran

Applies to

Product
Sytec 2000/3000/4000 and "i" Series and, Vertex PreSage, Luminex Vj Lj SLj Sytec 2000/3000/4000 and "i" Series and, Vertex PreSage, Luminex Vj Lj SLj Sytec 2000/3000/4000 and "i" Series and, Vertex PreSage, Luminex Vj Lj SLj Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
D3301T / D3302T Xray tube used on Sytec 2000/3000/4000 and "i" Series, Vertex PreSage, Luminex Vj Lj SLj, Pace, and Pace Plus See long description for more information. D3302T Xray tube used on Sytec 2000/3000/4000 and "i" Series, Vertex PreSage, Luminex Vj Lj SLj, Pace, and Pace Plus Part #: D3302T Part Type: GP TUBES Part Weight: 84.10 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: TUBE SYTEC AND SYTEC I FRU: Yes Status: OrderableComments: FLS - NEW 1994 TUBE CAT. EXCHANGE. REPLACES D1193GR .9 X .7 FS SYTEC 2000, 3000, 4000 OR PACEAND I SERIES (GREEN TUBE PART NUMBER IS: D3301T) DIMS 27X37X28H (INCHES) *MAXIRAY VERSION OF THIS TUBE IS D3309T* AM SOURCE = MEXICO FLS - FE ASSIST MS6588 FEB 2002 D3302T = GLASS TUBE YMS. DUE TO HIGH LEVEL OF DOA TUBESPLEASE REFER TO THE FOLLOWING SERVICE NOTES SN 108, T8011 & IF YOU HAVE A KYOSAN GENERATORSEE T6005 TO HELP ISOLATE THE ISSUE. ALSO IF YOU HAVE ANY DOUBTS PLEASE ASK YOUR LOACL ZSE OR CALL THE ONLINE CENTER. MAXIRAY EQUIVALENT IS D3309T
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.