GE Healthcare · CTLightSpeed260007370260007419
LightSpeed 4.X - Octane
Symptom
Customer sees a Pop-up window indicating the DAS temperatures are low. This does not prevent scanning only a warning. Customer sees a Pop-up window indicating the DAS temperatures are low. This does not prevent scanning only a warning
Error message
Error: 260007370, DAS Temperature is LOW. See Long Description for complete error Mon Feb 20 07:43:59 2006 Error: 260007370 Host: DAS Process: DCB Firmware File: ctemptst.c Line: 490 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 07:43:59:965 Log Series: 32 DAS Temperature is LOW. Lowest temperature was seen is on Converter Board 31. Measured: 10.5 degrees C. Acceptable Range: 15.0 to 62.0 degrees C. See Following Error Message for Converter Card List Error: 260007419 Data Acquisition System Temperature Is Low. See the LD for the complete error code Error: 260007419 Data Acquisition System Temperature Is Low. Mon Feb 6 08:37:56 2006 Error: 260007419 Host: DAS Process: DCB Firmware File: ctemptst.c Line: 474 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 1277 Type: Axial Exception Level: Pri/Soft Time: 18:27:37:070 Log Series: 20 Data Acquisition System Temperature Is Low
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2355552 MDAS 16 Convertor Card ( replaces 2322994-2 ) Used In LightSpeed 16 Gantry with MDAS. 2355552 MDAS 16 Converter Card used in LS 4X with 16 slice MDAS (replaces 2322994-2) Part #: 2355552 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.25 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: SERVICE, MDAS16 CONV BD FRU: Yes Status: Orderable Comments: Please consider ordering the used part when it is appropriate for your customer. It will save you substantial money. This is an MDAS16 CONV BD .compared to the SDAS. Make sure the correct Convertor is Ordered and replaced.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.