GE Healthcare · CTLightSpeedVCT200002352230011034

"Scan hardware Stopped Scan" with scan data errors.

LightSpeed 5.X (60kW)

Symptom

"Scan hardware Stopped Scan" with scan data errors. "Scan hardware Stopped Scan" with scan data errors

Error message

Error: 200002352 A HOST_SHUTDOWN was initiatied by the death of scanfilemgr. See long description for more details Mon Sep 26 06:56:58 2005 Error: 200002352 Host: chqxi Process: cupMonitor File: ssProcStop.c Line: 766 A HOST_SHUTDOWN was initiatied by the death of scanfilemgr

Error: 230011034 Unable to initialize scan data disk handling Thu Sep 4 12:04:35 2003 Error: 230011034 Host: imcct Process: scanfilemgr File: sdio_xfs_io_ops.c Line: 1228 Function: System Control : Scan Database Manager Scan Type: Not scanning Scan: 0/0/0 Unable to initialize scan data disk handling. Unable to create information file. Info file: /raw_data/info_file Mode: c2h Permissions: 667 (octal) Unix system error : 9 --- Bad file number.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution