McKesson PACS DICOM Modality Worklist request acquisitions, network data transfer from GenieAcq intermittently fails
Symptom
McKesson PACS DICOM Modality Worklist request acquisitions, network data transfer from GenieAcq intermittently fails McKesson PACS Dicom Modality Worklist request acquisitions, network data transfer from GenieAcq to remote DICOM host (i.e., Genie P&R, eNTEGRA, Xeleris, etc.) intermittently fails. The patient demographic data for the problem studies that fail to transfer is supplied via DICOM modality worklist query from a McKesson PACS. The only way to get the images contained in the problem study to transfer is to copy them from the problem study to another study and manually send.
Error message
Translation failed, status=-1, ErrCode=-102: Group 0x0020 Element 0x000d 2009/08/26 15:00:06 EST FRNUCLGEN01:ni;128:send.c[$Revision: 1.11 $] XferSendPatDataHdr() line 609 error 0x0EB0000F level 3 severity 3 {} "Send to FRNUCLXE01: Pat=48 Std=123 Ser=1 Is=1, Ty=STATIC D=2 Fr=1 Tr=0 Cc=0 LEHR_F" 2009/08/26 15:00:06 EST FRNUCLGEN01:ni;128:dicom/DicomXlator/GTODXlator.c[$Revision: 1.3 $] GEnieToDICOMXlator() line 357 error 0x0EB00039 level 3 severity 3 {} "Translation failed, status=-1, ErrCode=-102: Group 0x0020 Element 0x000d" 2009/08/26 15:00:06 EST FRNUCLGEN01:ni;128:dicom/DicomXlator/GTODXlator.c[$Revision: 1.3 $] GEnieToDICOMXlator() line 359 error 0x0EB0003D level 3 severity 3 {} "mfm error: bad UID syntax" 2009/08/26 15:00:06 EST FRNUCLGEN01:ni;128:send.c[$Revision: 1.11 $] XferSendPatDataHdr() line 635 error 0x0EB00006 level 3 severity 3 {} "NspDicomTransSendPatDataToDicom() failed" 2009/08/26 15:00:06 EST FRNUCLGEN01:ni;128:send.c[$Revision: 1.11 $] NspDicomIssSendReq() line 330 error 0x0EB00010 level 3 severity 3 {} "XferSendPatData() failed with error 246415366"Applies to
- Models
- Millennium, Xeleris
- Product
- IBACQ/X IBACQ/X IBACQ/X MG/MC/Myosight Millennium MG/MC Millennium MyoSIGHT MPR/MPS Millennium MPR Millennium MPS
- Software
- Genieacq All Genieacq All Genieacq All
- Component
- *Software Software Software
- Area
- Nuclear Medicine
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare Nuclear Medicine issues
- How do I capture and print COR and Camera Uniformity (CUA) result screens on Genieacq, then send to a processing station for ACR
- Collision and E-stop lights are red on the back of the gantry and no movement with Hand Held Controller (HHC) or the genieacq.
- Service Note 43783_1 Feb/08 Infinia/Ventri/Xeleris 1.1 & 2
- Need part numbers for the Millennium MG/MC table E-Stop
- How do I Delete a file on Genieacq that the system thinks is still acquiring
- Xeleris 1.1 DVD-RAMs do not always load back on to the same system that created them