GE Healthcare · CTBrightSpeedHiSpeed26011824150000
BrightSpeed HPower or Delight HPower
Symptom
Cannot perform tube warmup. Cannot perform tube warmup
Error message
Error: 260118241 (80-0180) Rotation board communications problem. See LD for full error Wed Mar 2 16:58:01 2005 Error: 260118241 Host: orp Process: orp_subsys File: gen_exec.c Line: 2148 Wed Mar 2 16:57:59 2005 Host : Jedi Ermes # : 260118241 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2148 Function : No System Function Reported Scan Type : Axial Scan: 50000/1/3 (80-0180)Rotation board communication problem. PPC board communication Watch Dog with Rotation board popped up because it did not get reply from Rotation board. Z6,Z7= 0h,0h Error: 260118241 (80-0180) Rotation board communication problem Code 80 Hardware Errors Error code: 80-0180H Message/explanation: Rotation board communication problem. PPC kV control board communication Watch Dog with Rotation board popped up because it did not get reply from Rotation board. Potential cause: Rotation board Control bus cable PPC kV control board Recommended action/ Troubleshooting guide: Verify the CAN cable between PPC kV control and auxiliary module is correctly connected to the Rotation board Check that rotation firmware is running (DS5 on rotation board blinking). If firmware OK, replace PPC kV control board.
Applies to
- Models
- BrightSpeed, HiSpeed
- Product
- BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5392132 Programmed Rotation Board - replaces 2409043, 2242440-11, 2214605-3 in Jedi Generator Auxiliary Module (This rotation brd. used on ALL HPower Systems and YMS Hispeed) See Long Description for Details. 5392132 Programmed Rotation Board - replaces 2409043, 2242440-11, 2214605-3 in Jedi Generator Auxiliary Module (This rotation brd. used on ALL HPower Systems and YMS Hispeed) Part #: 5392132 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: FRU- Programmed Rotation V3 board FRU: Yes Status: Orderable Comments: ** FE Assist ** For P500D or Xray RAD, order part number 2409043. CT and Vascular continue to order 5392132. 212010438 3/1/11 17Aug2010 - Promoted for ECO 2104218 212006824 for 305024364. Part #: 2409043 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: FRU-PROGRAMMED ROTATION BOARD FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution