GE Healthcare · CTLightSpeedVCT18220870636654301049466531
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE
Symptom
Cradle will move a few inches and stop. Cradle will move in a few inches and stop
Error message
Error: 182208 Alignment error between Encoder and Potentiometer. Potentioment Position 70636 (.01 mm) Encoder Position 65430 (.01 mm) Tolerance 2540 (.01 mm) See the LD for the complete error code Error: 182208 Alignment error between Encoder and Potentiometer. Tue Aug 5 15:05:36 1997 Suite: PCT2 Host: ETC Proc:Cradle Error: 182208 File: cmcEncoderRef.c 1.8 8/27/ Method: No Method Line: 152 Function: PATIENT ANATOMY POSITIONING : Cradle Drive Scan: 10494/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 15:05:35:935 Log Series: 27 HW error: Encoder C Pulse Alignment error with current Potentiometer reading. Current Position 66531 (.01 mm). Encoder Alignment Minimum (1456) and Maximum (1658) Potientiometer Window based on Last Potentiometer Reading (1658).
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2269175-3 LONGITUDINAL ENCODER ASM (Used on HSA, CTi, H1 QXi, Plus, Ultra, 16 slice H4 and HiSpeed QXi.) replaces 2269175-2. See long description for details Part #: 2269175-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Longitudinal Encoder Assembly with Dynapar encoder FRU: Yes Status: Orderable Comments: Per ECR 2036924, 212008898, 12/4/2006. ** FE Assist ** If no stock or unable to order the 2269175-3 , order 2269175-2. (212001261 5/22/07) Part #: 2269175-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Longitudinal Encoder Asm. NOTE:This asm. has LCC encoder/tested and sensor gap is 0.06-0.08mm FRU: Yes Status: Orderable Comments: Per ECR 2036924, 212008898, 12/4/2006
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.