Symptom

Yellow oily liquid seen oozing out of the base of the gantry. See Long Description Yellow liqid does not smell of urine

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5124371 VCT Heat Exchanger with Dielectric oil flushed and filled. See Long Description for Details. Part #: 5124371 Part Type: GP CONSUMABLES Part Weight: 180.78 LBR Repairable: No Returnable: No Hazardous Material: N Description: Performix PRO VCT Heat Exchanger with Dielectric oil flushed and filled. FRU: Yes Status: Orderable 2324118-3 VCT-JH4 Auxiliary Equipped. Aux Box See Long Description for details Part#:2324118-3 Part Type: GP CONSUMABLES Part Weight: 2.20 LBR Repairable: No Returnable: No Hazardous Material: N Description: VCT-JH4 AUXILIARY EQUIPED FRU: Yes Status: Orderable Comments: Set as Critical Part for Product CT_VCT, ms04278, 05Oct28 Added per VCT RSPL, ms13452 041205 Picture below 5105346-3 (replaces 5105346-2 and 5105346) RoHS Compliant Performix PRO 100 Smart Pump - VCT or HD See LD for details. 5105346-3 RoHS Compliant Performix PRO 100 Smart Pump - VCT or HD Part #: 5105346-3 Part Type: GP TUBES Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: RoHS Compliant Performix PRO 100 Smart Pump - VCT or HD FRU: Yes Status: Orderable Old FRU Compatible Yes Comments Rohs- Old part can be used on new systems. Flag to yes mar 18 2011 212008958 Per ECR 2100751, 212008898, 4 Jan 2011. Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable No Use on Hands Yes Use Demand History Yes Ultimate Prime 5105346-3 New FRU Backward Compatible Yes Old FRU Usable Yes Change Type Indirect ECO Number ECR2100751 Part #: 5105346-2 VCT PUMP / ADAPTER / HOSE ASM. Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Performix PRO 100 VCT Pump, Hose Asm. and Mounting Bracket Assembly, With furnished mounting hardware Pump Support Bracket and U-bolt Assembly and Flush and Filled with Dielectric Oil. FRU: Yes Status: Orderable Comments: This part is replacement for 5105346. Do NOT use 5105346 to replace a 5105346-2, ms13452 060131 D3194T PERFORMIX PRO-100 TUBE UNIT - Monopolar Hercules Tube used on 5.X (80/100), LS RT16, 7.X See Long Description for Details D3194T PERFORMIX PRO-100/VCT, RT16 and RT16 Xtra TUBE UNIT - Monopolar Hercules Tube used on 5.X (80/100), LS 5.X RT16, 7.X Part #: D3194T Part Type: GP TUBES Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: PERFORMIX PRO100/VCT (RG) & RT16 Systems FRU: Yes Status: Orderable Comments: Added per VCT RSPL, ms13452 041216 Comments: Added per VCT RSPL, ms13452 041216 FE ASSIST MSxxxx August 2005 D3194T = VCT / PRO 100 TUBE. VCT 32 or VCT 64 7.x systems may ONLY use D3194T tubes. 5.x HP 100 and HP80 Gdas systems may use either D3192T or D3194T tubes. '*** 2ND OPINION NOTES *** Perform Rotor Control Diagnostics--If tube rotor fails to reach RUN mode in three ACCEL steps look for: High Drive output currents Low Drive output voltages and Low Drive output frequencies. This indicates a load problem with tube. With rotor running turn off Gantry 120V power at PDU. Listen for excessive noise and short deceleration time. A good rotor will coast for several minutes without power. NOTE TO FES AND DCOS TEAM: IF SELLING TO AN END USER AND NOT INSTALLED BY A GE FE, AN END USER CERTIFICATION LETTER MUST BE SIGNED BY THE CUSTOMER. UPON RECEIPT OF SIGNED LETTER, ORDER CAN BE PLACED. **TO OBTAIN COPY OF CERT. LTR. CONTACT DCOS TEAM AT 800-558-2040 OPTION 2. **ANY QUESTIONS, CONTACT LYNN WESTBROOK OR ROD MCCRIMMON. ******* NOTE: D3194T tubes now manufactured, have a one inch longer high voltage cable. The longer cable is necessary for all Wide Bore Gantries. ( If a very old D3194T tube is installed on a Wide Bore Gantry, the high voltage cable may be too short, and a new D3194T tube must be ordered ).
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues