GE Healthcare · CTLightSpeedVCT230023053260107705260107203
LightSpeed 5.X (80/100kW)
Symptom
"Scan hardware Stopped Scan" with Axial Errors. "Scan hardware Stopped Scan" with Axial Errors
Error message
Error: 230023053 Failed on Gantry Firmware Mon Feb 4 07:19:52 2008 Error: 230023053 Host: aafgect1 Process: dailyPrep Rx File: DailyPrepUtils.c Line: 5433 Failed on Gantry Firmware. Cannot continue. Error: 260107705 Could not send axial drive message in the allotted time. Axial serial link is down. Possible causes: Axial CAN jumper, Axial fuse, interconnects, CAN devices. See LD for complete error Error: 260107705 Could not send axial drive message in the allotted time. Axial serial link is down. Possible causes: Axial CAN jumper, Axial fuse, interconnects, CAN devices. Mon Mar 14 10:34:57 2005 Error: 260107705 Host: tgp Process: tgp_subsys File: scanbusDriver.c Line: 233 Mon Mar 14 10:34:57 2005 Host : Table Gantry Processor Ermes # : 260107705 Exception Class : Abort Severity : Secondary File : scanbusDriver.c Line# : 233 Function : AXIAL CONTROL : Axial Communications Scan Type : None/Unknown Scan: 0/0/0 Could not send axial drive message in the allotted time. Axial serial link is down. Possible causes: Axial CAN jumper, Axial fuse, interconnects, CAN devices. Retries used: 0 Error: 260107203 Axial Motor Control (AMC) logic could NOT access the AXIAL drive via the CAN network.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5311936 Axial Encoder Assembly, Hub-Mounted Wheel. Used on H2, HP, VCT, and HD750. (Replaces 5182284-2 /5182284 / 2369747-2 / 2369747 / 2328307) See LD for details 5311936 Axial Encoder Assembly, Hub-Mounted Wheel. Used on H2, HP and VCT. (Replaces 5182284-2 /5182284 / 2369747-2 / 2369747 / 2328307) Part #: 5311936 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Shipping collector for Axial Encoder Assembly. FRU: Yes Status: Orderable Comments: Per ECR 2058601, 212008898, 10/8/08
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.