Pular para o conteúdo principal
GE Healthcare · CTLightSpeed200002738

LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2

Symptom

Scanner failed to scan with Image Allocation errors See Long Description for error examples Scanner failed to scan with Error: 2021 mkdir failed /usr/g/sdc_image_pool Input/output error and Error: 210000434 IMAGE ALLOCATION has returned null linkedTuple or null linkedTuple->data.

Error message

Error: 200002738, Creation of hard link to Image file for autoview failed See Long Description for complete error Tue May 23 10:37:48 2006 Error: 200002738 Host: ct01 Process: imagecreate File: ic_imagecreate_req.c Line: 1361 Image : 2928/3/17 Recon Sequence Id : 201 Creation of hard link to Image file for autoview failed Autoview file path : Varies Source file path :Varies Open file name : Varies Directory stat status : 0 File stat status : 0 Errno : OS error : 2 --- No such file or directory

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2 This is HP CPU host running LINUX with Pegasus IG. 308L.8_H3.3M4 ~ 4 or 8 slice MDAS. ~ Could be an Ultra, Plus or QXi system. Must run swokinstall -p to verify. ~ See Long description for details . LightSpeed 3.X MDAS4 Gantry and GOC2 Gantry - 2281177-2, DAS - 2272768, Console – 2341104-2, Table - 2276776, PDU - 2269902 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
308L.8_H3.3M4 for LightSpeed 3.X (Ultra, Plus, QX/I with Console GOC2 (w/Linux PC) Assembly (2315653) or Xtream (GRE) Console - showprods returns 308L.8_H3.3M4 for LightSpeed 3.X (Ultra, Plus, QX/I with Console GOC2 (w/Linux PC) Assembly (2315653) or Xtream (GRE) Console - showprods returns 308L.8_H3.3M4 for LightSpeed 3.X (Ultra, Plus, QX/I with Console GOC2 (w/Linux PC) Assembly (2315653) or Xtream (GRE) Console
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.