GE Healthcare · CTLightSpeed26012101326694
LightSpeed 5.X (80/100kW)
Symptom
Gantry angle number stuck at (some value) Gantry angle number stuck at S26.5
Error message
Error: 260121013 Gantry velocity less than the minimum allowed velocity. See Long Description for more details Wed Mar 29 11:28:24 2006 Error: 260121013 Host : Table Gantry Processor Ermes # : 260121013 Exception Class : Secondary Severity : Pri/Soft File : tilt mgr.c Line# : 1171 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Gantry velocity less than the minimum allowed velocity. Gantry velocity:-2 mdeg/sec Tilt position:-26694 mdeg Minimum allowable velocity:300 mdeg/sec
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2348298 GANTRY TILT TIMING BELT, 2.03 PITCH X 6 WIDE -used on LightSpeed 5.X and 7.X see LD for details P/N 2348298 GANTRY TILT TIMING BELT, 2.03 PITCH X 6 WIDE -used on LightSpeed 5.X and 7.X picture below Part #: 2348298 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: TIMING BELT, 2.03 PITCH X 6 WIDE FRU: Yes Status: Orderable Parts Offered as: Current: 2348298 Harvested: Not Available Exchange: Not Available Comments: FLS - TIMING BELT. ADDED PER THE HPOWER RSPL. EXPECTED M3 FOR THIS PROGRAM: FW19. MS20544 ON 3/24/03
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.