GE Healthcare · CTLightSpeedVCT26000741620053230023048
LightSpeed 5.X (60kW)
Symptom
Customer sees "Detector Temperature Is High. Image quality may be affected". Customer sees Detector Temperature Is High. Image quality may be affected
Error message
Error: 260007416 Detector Temperature Is High. Image quality may be affected See Long Description Tue Sep 6 09:07:45 2005 Error: 260007416 Host: orp Process: orp_subsys File: dtemptst.c Line: 364 Tue Sep 6 09:07:44 2005 Host : Das Ermes # : 260007416 Exception Class : Soft Severity : Sec/Soft File : dtemptst.c Line# : 364 Function : No System Function Reported Scan Type : Scout Scan: 372/1/1 Detector Temperature Is High. Image quality may be affected. Current Reading:40.41 Deg C A/D Counts:20053 Error: 230023048 Detector may not be at proper operating temperature to run Fast Cal See Long Description for complete error Wed Jan 18 07:25:50 2006 Error: 230023048 Host: ct99 Process: dailyPrepRx File: DailyPrepShellMgr.c Line: 2159 Detector may not be at proper operating temperature to run Fast Cal.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5340738 (replaces 2359114) HPOWER GANTRY FAN AND HARNESS used on 5.x gantry See Long Description for details This fan for BrightSpeed, HP100, HP60 and RT Systems. Part #: 5340738 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: GANTRY FAN AND HARNESS FRU: Yes Status: Orderable Comments: Per ECR 2091297, 212008898, 10/26/09. Part #: 2359114 <-- All GPRS stock is gone and this part is now obsolete Part Type: GP CONSUMABLES Part Weight: 8.82 LBR Repairable: No Returnable: No Hazardous Material: N Description: HPOWER GANTRY FAN AND HARNESS FRU: Yes Status: Orderable Comments: FLS - HPOWER GANTRY FAN AND HARNESS.ADDED PER THE HPOWER RSPL. EXPECTED M3 FOR THIS PROGRAM: FW19, 2003. MS20544 2/28/03.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required