GE Healthcare · CTBrightSpeedHiSpeed26011821150018
BrightSpeed HPower or Delight HPower
Symptom
Scanner Hardware Fault with high voltage spits. Scanner Hardware Fault with high voltage spits
Error message
Error: 260118211 ( 60-0313 )Inverter max. ILR current detected. This is a hardware detection of maximum current in serial resonant circuit. Z6,Z7= 0h,0h See long description for complete error Mon Feb 13 17:22:21 2006 Error: 260118211 Host: orp Process: orp_subsys File: gen_exec.c Line: 2150 Mon Feb 13 17:22:20 2006 Host : Jedi Ermes # : 260118211 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2150 Function : No System Function Reported Scan Type : Static Scan: 50018/1/4 (60-0313)Inverter max. ILR current detected. This is a hardware detection of maximum current in serial resonant circuit.
Applies to
- Models
- BrightSpeed, HiSpeed
- Product
- BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2266521-5 HV Tank (replaces 2266521-4 and 2266521-3 - backward compatable) used on LS 7.X, LS 5.X and LightSpeed RT. See Long Description for details 2266521-5 HV Tank (replaces 2266521-4 and 2266521-3 - backward compatable) used on LS 7.X, LS 5.X and LightSpeed RT. Dash Tanks includes the EMI Shielding needed for G-Das Part #: 2266521-5 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: JEDI H4 MPHVTANK WITH SHIELDS FRU: Yes Status: Orderable Part #: 2266521-4 Part Type: GP CONSUMABLES Part Weight: -5499945.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: JEDI H4 MPHVTANK WITH SHIELDS FRU: Yes Status: Orderable Part #: 2266521-3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 200.84 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: JEDI H4 MPHVTANK WITH SHIELDS FRU: Yes Status: Orderable Comments: added per ecr 2005924, as replacement for 2266521 and 266521-2. ms13452 040923 **If there is no Local Stock for the -3 part, you may order the -4 version which is backward compatible** ms0262 060815 *** 2ND OPINION NOTES *** Run the No-load KV test to verify HV tank is the error.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required