GE Healthcare · CTLightSpeed26000739111401125991140012600
LightSpeed 4.X - Octane
Symptom
"Scan hardware Stopped Scan" with DAS power errors. "Scan hardware Stopped Scan" with DAS power errors
Error message
Error: 260007391 The Data Acquisition System Has Detected An Error In The -/+12V analog DAS Power Supply See LD for more details Tue Aug 8 07:49:05 2006 Error: 260007391 Host: DAS Process: DCB Firmware File: pwrtst.c Line: 90 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 07:49:05:835 Log Series: 604 The Data Acquisition System Has Detected An Error In The -12V Analog DAS Power Supply.^B Measured: 0 mV , Expected range: -11401 to -12599 mV Tue Aug 8 07:49:05 2006 Error: 260007391 Host: DAS Process: DCB Firmware File: pwrtst.c Line: 90 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 0/0/0 Type: None/Unknown Exception Level: Sec/Soft Time: 07:49:05:735 Log Series: 604 The Data Acquisition System Has Detected An Error In The +12V Analog DAS Power Supply.^B Measured: 0 mV , Expected range: 11400 to 12600 mV
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2334456 12 volt power supply Asm - used on MDAS16 See Long Description Part #: 2334456 Part Type: GP CONSUMABLES Part Weight: 6.17 LBR Repairable: No Returnable: No Hazardous Material: N Description: PWR SPLY ASM, 12V DC (H16) FRU: Yes Status: Orderable Comments: FLS - PWR SPLY ASM, 12V DC (H16). ADDED PER H16 PROGRAM BY MS20544 ON 7/25/02.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required