Symptom

Scanner Hardware Stopped Scan with High Voltage errors

Error message

Error: 183296 Monitoring: (Measured 245.5 mA vs Ref 240 mA) See Long Description for the full error Error: 183296 Suite: CT01 Host: OBC Process: Generator File: gen_tech.c 1.36 Method: No Method Line: 529 Function: No System Function Reported Scan: 65001/8/38 Type: Axial Exception Level: Sec/Soft Time: 03:11:20:495 Log Series: 49 Selected Technic: 100 kV, 240 mA, 0 % Mod, Backup Timer 4324. VarmA: (Index: Start 0, Skip 0, Last 0) Last Cmd 240 mA. Monitoring: (Measured 245.5 mA vs Ref 240 mA) . 5 Total Samples, Tube Spit Ctr 0. (Rotor 0x0000ae04, Gen 0x9c19b5ff, Mon 0x00020000

Error: 183183 kV out of 3 % tol See Long Description for error example Error: 183183 Suite: CT01 Host: OBC Proc: Generator File: gen tech.c 1.32 Method: No Method Line: 791 Function: XRAY GENERATION : KV Loop 960822 1230 10922 Scan: 0/0/3 Type: Static Exception Level: Sec/Soft Time: 12:17:46:610 Log Series: 529 kV out of 3 % tol Selected technic: 100 kV, 50 mA Tot. samples = 3 Tot. kV samples above tol. = 0, below tol. = 1 Avg. of kV samples above tol. = 0.000 kV, below tol. = 53.871 kV 960822 1230 10922 Avg. of tot. kV = 84.396 kV

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.