GE Healthcare · CTHiSpeed
HiSpeed QX/i - Octane
Symptom
FE replaced image disk drive, but system image space is still not correct. See long description Check for the following issues with the system: 1. Image Works may not list exams properly. 2. Number of 512 image spaces is not correct. 3. df –h command shows an Image Pool is missing.
Error message
From SYSLOG: dksc0d1s0: Device not ready : Not ready to perform command see long description for details Open unix shell > login as root > cd /usr/adm > more SYSLOG IRIS 13# more SYSLOG Oct 10 08:58:42 6A:IRIS unix: dksc0d1s0 : <6> Device not ready <6>: Not ready to perform command (asc=0x4, asq=0x2)<6>
dksc0d2vol: 6 Device spin up failed, unable to use device -- corrective action necessary See long description for details Open unix shell and login as root cd /usr/adm more SYSLOG Use space bar to page through log and look for disk drive errors Apr 23 14:00:59 6A:bdicnd ls16 unix: dksc0d2s0: 6 retries exhausted Apr 23 14:00:59 1A:bdicnd ls16 unix: ALERT: I/O error in filesystem ("/usr/g/sdc image pool2") meta-data dev 0xee block 0x800040 Apr 23 14:50:27 6A:bdicnd ls16 unix: dksc0d2vol: 6 Device not ready 6 : Not ready to perform command (asc=0x4, asq=0x0) 6 Apr 23 14:50:27 6A:bdicnd ls16 unix: dksc0d2vol: 6 Device spin up failed, unable to use device -- corrective action necessaryApplies to
- Models
- HiSpeed
- Product
- HiSpeed QX/i - Octane Always MDAS4 101I.2_HSQXIM5 101.4_HSQXIM4 HiSpeed QX/i - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 101I.2_HSQXIM5 101.4_HSQXIM4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 101I.2_HSQXIM5 See LD for ClassM & C 101I.2_HSQXIM5 used on HiSpeed (YMS) only Class M software - 2395173 or 2401759 (6 pack) Class C software - 2395172 or 2401759 (6 pack) Patch G.6 Mandatory FMI# 25363 (P/N 5136941)
- Area
- CT
Full cause & step-by-step resolution