GE Healthcare · CTBrightSpeedVCT260107705260107203260107223260107460
BrightSpeed HPower or Delight HPower
Symptom
"Scan hardware Stopped Scan" with Axial Errors. "Scan hardware Stopped Scan" with Axial Errors
Error message
Error: 260107705 Could not send axial drive message in the allotted time. Axial serial link is down. Possible causes: Axial CAN jumper, Axial fuse, interconnects, CAN devices. See LD for complete error Error: 260107705 Could not send axial drive message in the allotted time. Axial serial link is down. Possible causes: Axial CAN jumper, Axial fuse, interconnects, CAN devices. Mon Mar 14 10:34:57 2005 Error: 260107705 Host: tgp Process: tgp_subsys File: scanbusDriver.c Line: 233 Mon Mar 14 10:34:57 2005 Host : Table Gantry Processor Ermes # : 260107705 Exception Class : Abort Severity : Secondary File : scanbusDriver.c Line# : 233 Function : AXIAL CONTROL : Axial Communications Scan Type : None/Unknown Scan: 0/0/0 Could not send axial drive message in the allotted time. Axial serial link is down. Possible causes: Axial CAN jumper, Axial fuse, interconnects, CAN devices. Retries used: 0 Error: 260107203 Axial Motor Control (AMC) logic could NOT access the AXIAL drive via the CAN network. See Long Description for complete error Fri May 19 19:13:11 2006 Error: 260107203 tgp tgp_subsys amc_mon.c 923 Running diagnostic: Axial2 Functional Fri May 19 19:13:11 2006 Host : Table Gantry Proce
Applies to
- Models
- BrightSpeed, VCT
- Product
- BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5311936 Axial Encoder Assembly, Hub-Mounted Wheel. Used on H2, HP, VCT, and HD750. (Replaces 5182284-2 /5182284 / 2369747-2 / 2369747 / 2328307) See LD for details 5311936 Axial Encoder Assembly, Hub-Mounted Wheel. Used on H2, HP and VCT. (Replaces 5182284-2 /5182284 / 2369747-2 / 2369747 / 2328307) Part #: 5311936 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Shipping collector for Axial Encoder Assembly. FRU: Yes Status: Orderable Comments: Per ECR 2058601, 212008898, 10/8/08
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required