Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26000745750084

LightSpeed 5.X (80/100kW)

Symptom

"Scan hardware Stopped Scan" with beam tracking errors. "Scan hardware Stopped Scan" with beam tracking errors

Error message

Error: 260007457 Beam Tracking Control Loop Error Exceeded The ABORT Tolerance See Long Description for complete error Wed Apr 27 18:54:40 2005 Error: 260007457 6 DAS DCB Firmware zaxis.c 564 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 50084/6/0 Type: None/Unknown Exception Level: Pri/Soft Time: 18:54:40:490 Log Series: 936 Beam Tracking Control Loop Error Exceeded The ABORT Tolerance. Tolerance : 400 umd Control Error : 448 umd Beam Side : B Side

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
D3634T (aka 2378211-2 ) Varian PERFORMIX PRO TUBE used on : HP80 AND LS RT4 See Long Description for details Part #: D3634T Part Type: GP TUBES Part Weight: 176.37 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: PERFORMIX PRO (VARIAN) TUBE UNIT FRU: Yes Status: Orderable Comments: FLS 11/19/04 (Shawn Fox 212001261) PLEASE remember to complete the Tube report for the tube you are replacing and return it with old tube. The blank Tube report comes with the new tube. If you have any questions, please contact Yolanda Scheper 262-548-2864. If you do not return tube report with old tube, warranty credit can be missed or delayed for your Zone
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues