Pular para o conteúdo principal
GE Healthcare · X-ray R&FOptimaDiscovery109045110904531000243

System displays a vertical homing error or “AutoPosition is not ready yet” message.

Symptom

System displays a vertical homing error or “AutoPosition is not ready yet” message. See Long Description System displays a vertical homing error or message that “AutoPosition is not ready yet.” In the message log there will also be a message “The OTS is not responding. Retry or reset the system.”

Error message

1090451 OTS Homing Failure. Vertical axis homing failure. Vertical encoder position is outside calibrated limits See Long Description 1090451 Customer Text: OTS Homing Failure Detailed Text: Vertical axis homing failure. Vertical encoder position is outside calibrated limits

1090453 An OTS CAN Command was not processed successfully. ThunderOTSDevice::_setVertAmpLock() failed See Long Descripton 1090453 Customer Text: An OTS CAN Command was not processed successfully Detailed Text: Retry or reset the system, ThunderOTSDevice::_setVertAmpLock() failed

1000243 AutoPosition is not ready yet. OTS homing has not been completed See Long Description 1000243 Customer Text: AutoPosition is not ready yet. Detailed Text: Retry in a few seconds or reset the system, OTS homing has not been completed

Applies to

Models
Optima, Discovery
Product
Optima XR640 Discovery XR650
Software
XR650 and XR640 M3 See Long Description for Config.txt file. atlasidc-2.1-6 atlas_linux_release-7.9-0 calypso-8.70-2 can_pci266_2615_smp-7.9-ctt627 CSD-2.4A-1 detector-1.1-6 dmcots-1.1-15 dm_Platform_release-1.1-15 ea3-3.2-5 enggui-15.4-3 FlightRecorder-15.13-ATL.7.9 GlobalOneChinaFW_V4-1-2 IIPClassA-3.3.12-3 IIPClassM-3.3.12-3 IPCots-4.2-1 iui_globalone_release-5.1-23 j2re-1_4_2_13-linux-i586 j2sdk-1_4_2_13-linux-i586 jedi_lightning_magic-6.33BF6-1 machineDB-2-5 magic-aux_tools-1-1 magic-config_prod_GlobalOne-1-1 magic-cots-4-58 mcf_release-1.7-0 pnf-2.5-3 pos_download_mke-1-22 psys_can_pci266_2615_smp-7.9-ctt627 rcim2_smp_linux-7.9-0 RCOC-1.2.8-0 RTIP_G1_Magic_Release-FW17-RC3 saber_release-2.3.64C-1 SysConfig_Linux-103B-ATL.7.8.6 tao_linux_release-7.9-0 CD-R Label Data: 1) Business Name: Global Diagnostic X-Ray 2) CD-R Part Number: 5316167-3 3) Revision: 4 4) Software Title: Application Software 5) Disc Number: 1 of 1 6) Product Name: Discovery XR650/Optima XR640 7) Legally Required Messages ONLY: 8) Duplication Date: 9) Copyright Date: 2009
Component
OTS
Area
X-ray R&F

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required