Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT23000089750074

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

"Scan hardware Stopped Scan" with scan data errors. "Scan hardware Stopped Scan" with scan data errors

Error message

Error: 230000897 Write to scan data disk failed. See LD for more details Tue May 23 07:22:28 2006 Error: 230000897 Host: darc Process: dataacq File: raw_file.cxx Line: 250 Function: Data Acquisition : Scan Data Save Scan Type: Not defined Scan: 50074/1/1 Scan Seq Id: 0 Write to scan data disk failed. Status: -1 Most Recent System Error: OS error : 98 --- Address already in use

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5248406 (replaces 5116147 and 5116147-2) SCSI cables between SDDA (JBOD) and VDARC See LD for Parts Details SCSI Cables between the SDDA (JBOD) and VDARC It is suggested that both cables be ordered and replaced at the same time. Part #: 5248406 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: CABLE, SCSI 900mm, VHDCI 68m - VHDCI 68m, with Tin plated Copper Tape FRU: YesStatus: Orderable Comments:Per ECR 2049714, 212008898, 4/7/08 Old FRU Compatible Yes Comments Per ECR 2049714, 212008898, 7/23/08. Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable No Use on Hands Yes Use Demand History Yes Ultimate Prime 5248406 New FRU Backward Compatible Yes Old FRU Usable Yes Change Type Indirect ECO Number ECR2049714 Part #: 5116147 Part Type: GP CONSUMABLES Part Weight: 2.20 LBR Repairable: No Returnable: No Hazardous Material: N Description: CABLE, SCSI VHDCI68m-VHDCI68m x 610mm FRU: Yes Status: Orderable Current: 5116147 Harvested: Not Available Exchange: Not Available Comments: Added per VCT RSPL post M3, ms13452 050426 Part #: 5116147-2 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: CABLE, SCSI VHDCI68m-VHDCI68m x 610mm -B- FRU: Yes Status: Orderable Current: 5116147-2 Harvested: Not Available Exchange: Not Available Comments: Added post M3 to VCT RSPL, ms13452 050426
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.