Pular para o conteúdo principal
GE Healthcare · CTLightSpeed23000090053852633554432123321812672703175000

LightSpeed 5.X (80/100kW)

Symptom

Scanning aborted with Buffer overrun detected during scan data transfer

Error message

Error: 230000900 WARNING slow write to disk See llong decsriptions for full error Thu Jan 29 08:54:34 2009 Error: 230000900 Host: darc Process: dataacq File: file_save_thread.cxx Line: 194 WARNING slow write to disk: 51.538526 sec to write 33554432 bytes; scan CT07/152/3/3-1233218126/727031; scan seq 75000 Example: This took 51 seconds to write 33 mb. which is extremely long. 51.538526 sec to write 33554432 bytes;

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5114772 Darc2 Computer Dual Disk used ONLY on 16 slice MDAS and 4/8/16 slice GDAS See LD for more info 5114772 Darc2 Computer Dual Disk used ONLY on 16 slice MDAS and 4/8/16 slice GDAS Part #: 5114772 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: DARC Computer - Dual Disk H16 MDAS, GDAS FRU: Yes Comments: Added per GOC4 RSPL Replace the DARC2: 1) If you get a Westville DARC P/N 5114772 load the DARC software to recover. 2) If you get Jarrell p/n 5147442 perform LFC using 06MW29.7 software.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required