Pular para o conteúdo principal
GE Healthcare · CTLightSpeed260115245237551771026011518512727441349271

LightSpeed 5.X (60kW)

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260115245 MTCB cradle encoder delta data out of range. See long description for full sample error. Thu Aug 7 08:34:27 2008 Error: 260115245 Host: tgp Process: tgp subsys File: mtcb cradle monitor.c Line: 285 Thu Aug 7 08:34:27 2008 Host : Table Sub-System Ermes # : 260115245 Exception Class : Secondary Severity : Pri/Soft File : mtcb cradle monitor.c Line# : 285 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 MTCB cradle encoder delta data out of range. Possible Encoder, cabling or MTCB issue. cradle current position = 23755 cradle previous position = 17710 Pacsci drive command position = -96 position offset = -396 previous position offset = -396

Error: 260115185 Cradle velocity is out of tolerance Error. 260115185 Cradle velocity is out of tolerance cradle_position(um) : 1272744 target_position(um) : 1349271 velocity (um/s) : 75000 fpga status : 0x8104

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2342727-4 : 5.X Cradle encoder. This part replaces the 2342727-3 and 2342727-2 This is an absolute position encoder with no pot. See long description for details Displaying Change-Up Data: Part "2342727-2" and "2342727-3" changed to 2342727-4 Part #: 2342727-4 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: FRABA OCD H-Power String Encoder FRU: Yes Status: Orderable Comments: Added by ecr: 2020264. Please continue to use stock of 2342727-3, until depleted. -3 and -4 are mfg alts of one another. ms13452 Revised per to Asset confirmation Comments: Added per ECR2011883 to replaced the 2342727-2 The 5.x table design incorporates Absolute encoders used for cradle and elevation, instead of incremental encoders. The absolute encoder has the capability to generate a unique binary 'word' for each encoder shaft position, allowing the encoder to provide positional information instantly upon power-up, unlike incremental encoders. Therefore, potentiometers are NOT needed for cradle or elevation encoder circuits.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.