GE Healthcare · CTLightSpeed18222049439422651822071049470636
LightSpeed 4.X - Octane
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 182220 Cradle drive is NOT stopping within 1520 msec. See long description for full error Mon May 12 2008 01:43:49 Error: 182220 Host: ETC Process: Cradle File: cmcShutDown.c 1.54 2/5/96 Line: 100 Function: PATIENT ANATOMY POSITIONING : Cradle Drive Scan: 49439/2/0 Type: None/Unknown Exception Level: Pri/Soft Time: 20:22:14:155 Log Series: 709 HW/SW error: Cradle drive is NOT stopping within 1520 msec. Current Position 42265 (.01 mm) Cradle Direction: 1 (0 = Out 1 = In [Gantry]). flags:0x300e97 Error: 182207 HW error: Alignment error between Encoder and Potentiometer See Long Description for complete error Tue Aug 5 15:05:39 1997 Error: 182207 Host: ETC Proc:Cradle File: cmcCorrErr.c 1.22 3/11/92 Line: 51 Function: PATIENT ANATOMY POSITIONING : Cradle Drive Scan: 10494/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 15:05:39:320 Log Series: 28 HW error: Alignment error between Encoder and Potentiometer. Potentioment Position 70636 (.01 mm) Encoder Position 65430 (.01 mm) Tolerance 2540 (.01 mm)
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2269175-3 LONGITUDINAL ENCODER ASM (Used on HSA, CTi, H1 QXi, Plus, Ultra, 16 slice H4 and HiSpeed QXi.) replaces 2269175-2. See long description for details Part #: 2269175-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Longitudinal Encoder Assembly with Dynapar encoder FRU: Yes Status: Orderable Comments: Per ECR 2036924, 212008898, 12/4/2006. ** FE Assist ** If no stock or unable to order the 2269175-3 , order 2269175-2. (212001261 5/22/07) Part #: 2269175-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Longitudinal Encoder Asm. NOTE:This asm. has LCC encoder/tested and sensor gap is 0.06-0.08mm FRU: Yes Status: Orderable Comments: Per ECR 2036924, 212008898, 12/4/2006
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required