Symptom
Image reconstruction failures. Image reconstruction failures
Error message
Error: 230005024 image recon seq id: 18997Rebin failed in row delay, status=2 See Long Description for complete error Mon Feb 6 12:12:50 2006 Error: 230005024 Host: ice-post Process: sdcctrl File: sdcctrl_events.cxx Line: 447 image recon seq id: 18997Rebin failed in row delay, status=2 Error: 230006529 Recon Mgr is sending queue ack to queuemgr with a RECON_JOB_FAILED_RETRY status See Long Description for Full Error Tue Feb 1 06:33:34 2005 Error: 230006529 Host: ice Process: reconmgr File: rm_recon_events.c Line: 329 Function: Image Generation : SBC Processing Job Type: Prospective Axial Scan: 25088/2/9 Image: 25088/3/25 Recon Seq Id: 957 Recon Mgr is sending queue ack to queuemgr with a RECON_JOB_FAILED_RETRY status Error: 230013015 xx DAS views of 100 retrieved from the scan database failed checksums See Long Description for great info on this error Problem: The Following errors all point to a scsi Scan data disk issue. Using the gesys_Ct01.log and iceConsole.log is a must, together both logs point to the Scan data disk chain. Check DIP stats if stats are clean the das data is good. The iceConsole log is a little difficult to read and hard to correlate to specific dates and tim
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2225004-10 73 GB destroked to 18GB 15000 Used on Octane OC with 9Gb OS hard drive. See Long Description for more detail. This part is used on Octane Operators Consoles that have a 9Gb OS hard drive and need a destroked Scan Data Disk. See SN0711061b for jumper settings. Part #: 2225004-10 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: 73 GB destroked to 18GB 15000 RPM Scan Data Drive FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.