revolution xrd 2x platform (lightning) - error 3601144 reported by Jedi generator 40-0106 : error msg from Rot…
revolution xrd 2x platform (lightning) - error 3601144 reported by Jedi generator 40-0106
Symptom
revolution xrd 2x platform (lightning) - error 3601144 reported by Jedi generator 40-0106 : error msg from Rotor FW; Phases error. This failure happens intermittently at prep (several imes per day).
Error message
3601144 - Generator error 40-0106;(4): error msg from Rotor FW; Phases error. The rotation board has detected that the current in the anode stator does not show the correct phase shift between main and auxiliary.
Applies to
- Models
- Revolution
- Product
- Revolution XRd
- Component
- jedi generator - anod starter - rotation board - capacitor
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray R&F issues
- ADx-E Advantx-E console displayed a 6C00 Error (Digital Imaging Subsystem error - Error status received in IDC Prep Reply) when the first exposure of detector bad pixel calibration was attempted; cali…
- Revolution XR/d After a Load From Cold, the AWS exhibits strange operation and the SUIF does not have any Positioner Configuration and Calibration selections due to the use of the incorrect Jumpstart…
- Revolution XRd2X Worklist query on "All Systems in this Modality" does not retrieve anything after FMI 10791A is completed due to the default query set to DX.
- Revolution XR/d error 4280 when trying to make exposure to table or wallstand detector
- Revolution XRd 2X The HP Workstation reboots itself.
- Def8000, Def5000, Revolution XRd and XRD2X Latent lead marker image burned into the detector.