GE Healthcare · CTLightSpeed178568
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane
Symptom
No flashing eights on gantry display and Scanning hardware reset failed
Error message
Error: 178568 Failed to establish socket connection to STC in 10 seconds See Long Description for full error Fri Dec 24 04:38:04 2004 Error: 178568 Host: ct02_sbc Process: fwmgr File: NBSSocket.c Line: 388 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failed to establish socket connection to STC in 10 seconds
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-296317P3 XENTEK POWER SUPPLY See Long Description for more information including stc and ETC power supply locations Part #: 46-296317P3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: POWER SUPPLY FRU: Yes Status: Orderable Exchange: Not Available Comments: FLS - XENTEK POWER SUPPLY U/O GANTRY & TABLE CTI AND GENESIS AND HORIZON USE P3'S ON CTI ONLY 10/31/96 MS08992 GVR ---------------------------------------------------------- 46-296317P3 REPLACED T3202GT /46-296317P2 PER PCN 189892. MS4636 12/2/96 ---------------------------------------------------------- ALSO USED ON QXI LIGHTSPEED TO BRASIL: IF NO STOCK, CHANGEORDER TO T3202JR. COLLECTION OF DEFECTIVES START7/23/01 MS5277 TAIYO YUDEN P/N 8230 +5VDC, +15VDC, -15VDC, +24VDC,IN 120VAC, 3.6A 50/60HZ MS5900 WAS T3202JR FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA FE ASSIST: IN ORDER TO INCREASE REVENUE GLOBAL PARTS HAS CREATED A PROGRAM TO HARVEST PARTS FROM RENEWABLE RESOURCES. THESE PARTS OFFER DIRECT SAVINGS TO THE BUSINESSAND THE FIELD. ORDER THE "U" PART IF AVAILABLE. IF NO "U" USED PARTS ARE AVAILABLE CARES WILL CHANGE THE ORDER TOA NEW PART.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.