Pular para o conteúdo principal
GE Healthcare · CTBrightSpeed260100040260100042

BrightSpeed HPower or Delight HPower

Symptom

Can't perform Scout Scan. Can't perform Scout Scan Please go to the error concept and check all that apply The customer may see Scanner Hardware Stopped Scan message in the operator console area 1 Use the following script and search the message log at time of Scout failure to get the correct errors. [insite@erct ~]$ /usr/g/insite/bin/viewlog 2 Type those errors in the psdb search box separate by spaces 3 Go to the error concept and check the appropriate errors

Error message

Error: 260100040 Timeout on Table gate signal (Cradle Sync Pulse). See LD for complete error Mon Nov 7 16:37:30 2005 Error: 260100040 Host: tgp Process: tgp_subsys File: scTrigSeqActions.c Line: 1716 Mon Nov 7 16:37:31 2005 Host : Table Gantry Processor Ermes # : 260100040 Exception Class : Abort Severity : Primary File : scTrigSeqActions.c Line# : 1716 Function : XRAY GENERATION : Table - Gantry Sync Scan Type : Scout Scan: 209/1/1 Timeout on Table gate signal (Cradle Sync Pulse).

Error: 260100042 Cradle was NOT ready See the long description for the full error code Tue May 17 16:02:03 2005 Error: 260100042 Host: tgp Process: tgp_subsys File: scHdwRdy.c Line: 70 Tue May 17 16:02:02 2005 Host : Table Gantry Processor Ermes # : 260100042 Exception Class : Soft Severity : Sec/Soft File : scHdwRdy.c Line# : 70 Function : No System Function Reported Scan Type : Scout Scan: 106/1/1 Cradle was NOT ready

Applies to

Models
BrightSpeed
Product
BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2342729-3 Elevation or Cradle Drive Amplifier used on the 5.X table - same driver amp is used for both See LD for additional part information Part #: 2342729-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: H-POWER TABLE SERVO DRIVE, EMC EDITION 2 COMPLIANT FRU: Yes Status: Orderable Comments: **Please consider ordering the used part when it is appropriate for your customer. It will save you substantial money.** Part added per DST-HP60 16 slice RSPL. 040906 MS13452.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues